Setting The Setpoints - Siemens SIPROCESS GA700 Quick Start Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

7.10.2.2

Setting the setpoints

Procedure
1. Main menu > "2. Settings" > "08. Calibrations" > "1. Setpoints, tolerances, purge
2. Enter zero gas setpoints: .
3. Enter span gas setpoints:
Quick Start
Compact Operating Instructions, 01/2013, A5E31805656-01
time" > "1. Setpoints"
In this component-specific menu you can enter the setpoints for zero gas and span gas
calibrations for each measuring range.
The unit and format of the input correspond to the default settings of the respective
component.
Structure of the navigation lines
Zero gas measuring range 1 to 4
Span gas measuring range 1 to 4
Reduced display. One line or input of a setpoint is possible for each measuring range.
1)
With OXYMAT 7 always in percentage (0 to 100%).
2)
Note
Setting ranges
The setting ranges of the setpoints (possible minimum/maximum values) are limited by
the assigned start-of-scale values and full-scale values.
"Zero gas range 1" to "Zero gas range 4" parameter fields.
– Authorize access with Expert PIN, if necessary.
– The assignment of one setpoint is sufficient for a total calibration.
– Zero gas setpoint < span gas setpoint
"Span gas range 1" to "Span gas range 4" parameter fields.
– The assignment of one setpoint is sufficient for a total calibration.
– Span gas setpoint > zero gas setpoint
Recommendation: Span gas setpoint at 70 to 100% of the current measuring range.
1)
1)
Commissioning
7.10 [2.08] Settings > Calibrations
Setpoint
2)
Setpoint
2)
75