Commissioning
7.10 [2.08] Settings > Calibrations
5. Assign measuring range(s): "Measuring range" parameter field.
Note
Setting the measuring ranges
Check the settings of the measuring ranges before you assign a calibration to a range.
Make sure that the measuring ranges are set appropriately for the respective application.
Adjust the settings of the measuring ranges again, if necessary: [2.03] Settings >
Measuring ranges (Page 51) menu.
6. Start the wizard for a calibration or validation.
Rectifying errors
If the device aborts a calibration due to errors, an error code, e.g., "F0123", and a plain text
error message are displayed in the result window.
1. Write down the error code.
2. If you cannot rectify the cause of the error yourself, forward the error code to the service
80
If you have set "Total" for the scope in step 4, the selected measuring range acts as the
"leading measuring range".
– "Start validation" parameter field
A validation evaluates the current device status with respect to the last calibration. The
need for a new calibration can be derived from the validation results.
– "Start calibration" parameter field
Follow the instructions on the display. The success of the calibration/validation is
displayed in the result window once the process is complete.
technician in charge.
Compact Operating Instructions, 01/2013, A5E31805656-01
Quick Start