Setting the calibration tolerances

Procedure
1. Main menu > "2. Settings" > "08. Calibrations" > "1. Setpoints, tolerances, purge
2. Specify the zero point: "Zero point" parameter field.
3. Specify the span point: "Span point" parameter field.
Quick Start
Compact Operating Instructions, 01/2013, A5E31805656-01
time" > "3. Calibration tolerances"
The settings of this menu cause the device to monitor the deviations of the zero point or
span point compared to the last calibration. The reaction to a tolerance violation is
different for validation tolerance and calibration tolerance.
Note
Tolerance violation
If the factory-set tolerance value is exceeded, the device outputs a maintenance alarm.
The indicated measured values are no longer valid:
• Evaluate the message text of the maintenance alarm. Remedy the causes of errors
mentioned in it, for example, incorrectly set setpoints.
• If the cause of a tolerance violation cannot be traced back to incorrect operator inputs,
contact the service department and take the device out of service.
– Authorize access with Expert PIN, if necessary.
– Enter the zero point in %. The percentage refers to the value of the smallest span.
Enter the span point in %. The tolerance in the span point refers to the span or the full-
scale value of the selected measuring range.
Commissioning
7.10 [2.08] Settings > Calibrations