Commissioning
7.10 [2.08] Settings > Calibrations
7.10.2.5

Setting the sample gas purging time

Requirement
The sample gas purging time is the time that elapses from switching of the solenoid valve
until measurement of the sample gas. Because of the different use environments, it is
possible to assign the sample gas purging time.
Depending on the hardware of the device, the sample gas purging time can be set device-
specific or module-specific:
Table 7- 23
Device-specific
If only one analyzer module is installed or if a
common gas path exists for two analyzer
modules (series connection).
Check the setup of the gas path as well as the parameter assignment of the process tags
before you set the sample gas purging time: [2.12] Settings > Gas path/process tag label
(Page 69).
Procedure
1. Main menu > "2. Settings" > "08. Calibrations" > "1. Setpoints, tolerances, purge
2. Specify duration: "Sample gas purging time" parameter field.
78
Effect of the sample gas purging time
time" > "4. Sample gas purging time"
– Authorize access with Expert PIN, if necessary.
– Increase/reduce preset value.
– The device stays in "Function check" operating state for the assigned duration.
Module-specific
If two analyzer modules with separate gas paths
are installed, you must set the parameters
separately for each analyzer module.
Compact Operating Instructions, 01/2013, A5E31805656-01
Quick Start