7.6.2

Setting the measuring ranges

Procedure
Note
Component-specific setting
Measuring ranges are component-specific. Repeat the procedure described below for all
components, if necessary.
1. Component > Main menu > "2 Settings" > "03 Measuring ranges"
2. Open selected measuring range: Press <ENTER>.
3. Set up the measuring range:
4. Enable/disable measuring range: "Use measuring range" parameter field.
5. Set additional measuring ranges/parameters, e.g., Setting Autoranging (Page 54).
Quick Start
Compact Operating Instructions, 01/2013, A5E31805656-01
The menu includes a navigation line for setting of autoranging.
One navigation line is reserved for each of the four measuring ranges that can be set:
Structure of the navigation line for each measuring range
Navigation no.
On/Off: signals if the measuring range is currently active and in use.
1)
The parameter display of the measuring range is opened.
Refer to the specifications under Overview of measuring ranges (Page 51).
– Enter start of measuring range: "Start-of-scale value" parameter field
Authorize access with Standard PIN, if necessary.
– Enter end of measuring range: "Full-scale value" parameter field.
If you select the "On" setting, the measuring range is used for autoranging.
Measuring range start - measuring range end
7.6 [2.03] Settings > Measuring ranges
Activation status
Commissioning
1)
Edit
53