Commissioning
7.6 [2.03] Settings > Measuring ranges
7.6.3

Setting Autoranging

Procedure
Note
Autoranging
Autoranging is not possible if the measuring ranges were set incorrectly. In this case
measuring range 1 is selected as active measuring range.
Note
Component-specific setting
Measuring ranges are component-specific. Repeat the procedure described below for all
components, if necessary.
1. Main menu > "2 Settings" > "03 Measuring ranges"
2. Call autoranging: Navigation line "MR switchover"
3. Specify measuring range: "Measuring range selection" parameter field.
54
The menu includes four navigation lines for setting the measuring ranges and one
navigation line for setting autoranging.
The parameter display of autoranging has a parameter field and a display box.
– Authorize access with Standard PIN, if necessary.
– Select one of the offered measuring ranges or autoranging:
If the message "Active" is displayed in the "Autoranging" line, autoranging is activated.
If "Not possible" is displayed, correct the start and end values of the individual
measuring ranges: Setting the measuring ranges (Page 53).
Compact Operating Instructions, 01/2013, A5E31805656-01