Siemens SIPROCESS GA700 Quick Start Manual page 66

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Commissioning
7.8 [2.09.5] Digital outputs
Table 7- 14
Function
(pre-selection)
Limits
Table 7- 15
Function
(pre-selection)
Ext. warning limit
Table 7- 16
Function
(pre-selection)
Measuring ranges
66
Function (pre-selection) "Limit"
Purpose/resulting settings
Signal limit violation
1. "Component" parameter field:
Authorize access with corresponding PIN, if necessary.
Select available component.
2. "Function" parameter field:
Select one of the two assigned limits.
Function (pre-selection) "Ext. warning limit"
Purpose/resulting settings
Monitor with an external warning limit.
1. "Function" parameter field:
Authorize access with corresponding PIN, if necessary.
2. Set one of the eight assigned external warning limits.
Function (pre-selection) "Measuring ranges"
Purpose/resulting settings
Specify measuring range to be displayed.
1. "Component" parameter field:
Authorize access with corresponding PIN, if necessary.
Select available component.
2. "Function" parameter field:
Select one of the four available measuring ranges.
Compact Operating Instructions, 01/2013, A5E31805656-01
Quick Start