9 Troubleshooting

If any problem occurs during the operation, try following methods to recover.
The host device which is the access point may cause a problem. Refer to the
instruction manual of your host device.
9.1

The reader does not start up or cannot communicate with a PC.

Is power supply voltage within the specifications?
Power supply voltage should be DC 24V ± 10%.
If the DC voltage is not within the specified range, the reader may not work
properly and there is a possibility of damaging the reader.
Is the LP-ABR connected with the host device properly?
Make sure the LP-ABR is connected with the host device properly.
In RS-232C interface, all the communication conditions, such as baud rate must
be consistent.
Is the format of serial command correct?
Serial commands input are effective in stand-by state.
Add a carriage return code [CR] at the end of the serial command.
Push Enter key at the end to input the serial command by typing on Terminal
software.
Transmit the serial command "WSETS" to save the settings to internal flash
memory. If the reader is turned off the power without saving settings, the reader
will start up in the previous settings next time.
An interval of 500ms between each command will avoid unwanted error during
settings.