Troubleshooting
Description
Message displayed
showing ± 30 minutes
left of battery operation
Low Battery
Connect to charger
End Battery
End Battery
Connect to charger
Message displayed in the
TPN programs, that the
parameters entered are
incorrect and cannot be
calculated by the pump
Wrong TPN Date
Check Data
Pump is in lock-mode
Lock out
A fatal error is detected
Error XXX
Send for Service
VET-PRO Operator's Manual
Result
Possible Cause
If pump
Message will be displayed as
won't be
long as the internal battery
connected
voltage is less than 7.1v.
to mains,
battery will
be depleted
Infusion
Battery is depleted.
ceases
The pump cannot be used
before being connected to
mains.
Program
The set parameters are
will not
impossible
start
Volume or time parameters
operating
are incorrect
Program
The
program
parameters
second-level locking, which
can not be
enables daily operation of
changed
the set program, without the
ability to change parameters.
The pump will advise the
Infusion
user to send the pump for
ceases
service.
Connect to Mains
Connect to Mains and
wait 2 minutes before
operating.
Check data and change
accordingly.
to
perform.
is
set
in
No action required.
Restart the pump. If the
alarm does not stop
pack
properly, in its original
packaging, and send it
for service.
Remedy
the
pump
57