Info Mode

Pressing on the info key during operation will
display the total Volume To Be Infused and the
Infused Volume. The screen will be displayed
for 7 seconds.
Pressing twice on the info key during operation
will display the battery status.The screen will be
displayed for 7 seconds.
Pressing 3 times on the info key during
operation will display the administration set
temperature.The screen will be displayed for 7
seconds.
This option is not available in the 25 steps
program.
Pressing 3 times on the info key, during PCA
program, will display Boluses History, from
beginning of the program operation.
Pressing 4 times on the info key during
operation will display the time left until the end
of the program. This Option is not available in
the 25 steps program and in the PCA program.
Pressing 4 times on the info key during TPN
operation will display the program status using a
graph. The performed part of the trapeze will be
painted black. The screen will be displayed for 7
seconds.
Pressing 4 times on the info key during PCA
operation will display the data of the present
protocol operating.