Connection of Pump to IV Pole
1.
Connect the giving set to the medication bag. Mount the bag on an IV pole.
2.
Click the pump into the Charger. Mount the Charger on an IV pole.
3.
Connect the Charger via a standard power cord to a grounded AC outlet.
4.
Open the door of the pump and insert the administration set into the pump's
tubing guide (as described on page 17)
5.
Switch the pump ON using the ON/OFF key
(press until the self-test screen appears).
6.
The program and status will appear on the
display followed by the last rate setting.
7.
Flush the administration set by pressing the SYRINGE key, and confirm
with START/OK.
8.
Set the required program (see program select option on page 24) and continue
as described for every program setting.
9.
Verify that infusion is proceeding normally before leaving the pump
unattended.

Continuous Program

Switch the pump ON using the ON/OFF key.
Pump displays program name and status. The
status is either "Lock out" or "program"
(locking and unlocking the pump – see page
60).
VET-PRO Operator's Manual
C
CONTINUOUS
O
N
PROGRAM
T
Press NO to change
C
CONTINUOUS
O
N
PROGRAM
T Press NO to
C
CONTINUOUS
O
N
LOCK OUT
T
change
27