Use the STOP/NO key to erase the displayed drug name.Use the arrow keys
and keys of 4 and 6 to move between letters. Select the letters using the
START/OK key. Once the protocol name is completed, return the select key
to the empty space, before the letter A and repeatedly press the START/OK
key until the display returns to the Protocol List.
11.
Pump will display the rate screen (the chosen
protocol will be displayed on the left column of
the screen). Use the numeric keys to set a new
value (up to 100 ml/hr) or approve the previous
value, using the START/OK key. Rate can be
displayed in mg/ml, as well. Using this option
– go to step 12, if not, continue to step 13.
From the rate screen press the STOP/NO key 3
12.
times to display the concentration screen. Use
numeric keys to set concentration and approve
by pressing the
back to the protocols list.
To return from mg to ml, press STOP/NO key
twice, from rate screen, enter a concentration of
0 mg/ml, and press the START/OK key for
approval.
13. Pump will display the Bolus screen. Use the
numeric keys to set a new value or approve the
previous value, using the START/OK key.
14.
Lock time is the minimal time between
boluses. When a bolus is required during the
lock time, a beep will be sounded but no bolus
will be given. Use the numeric keys to set a
new value or approve the previous value, using
the START/OK key.
4 H Limit is the maximum volume to be given in 4
15.
hours. Once this volume will be infused the pump will
operate in the KVO mode, to complete the 4 hours.
Use the numeric keys to set a new value or approve the
previous value, using START/OK key.
16. PCA program operates only in a lock mode.
Press the START/OK to lock the program or
STOP/NO to go back to protocols list and
scroll again through the data.
40
START/OK. Pump will go
P
Rate 50 ml/h
Adjust & press OK
A
P
Concentration
C
A
0.5 mg/ml
P
BOLUS 5 ml
Adjust & press OK
A
P
LockTime 60
Adjust & press OK
A
P
4 H Limit 10 ml
Adjust & press OK
A
P
Lock:Press OK
C
Scroll:Press NO
A
VET-PRO Operator's Manual