Battery Operation - VET-PRO VIP 2000 Operator's Manual

Veterinary infusion pump
Table of Contents

Advertisement

Scroll through the setting parameters, using the
ARROW keys, until the "Maximal Rate "
option. Confirm with START/OK key.
Set the maximal rate, using the ARROW keys
and confirm with START/OK key.
F
Note:
F
Note:

Battery Operation

The VET-PRO can operate on battery power, enabling operation when the
patient is being moved or during electrical power failure. When the pump runs
on the battery power, the AC icon is off. At full charge, the battery provides 9
hours of operation at an infusion rate of 250ml/hr.
I
Warning
When the pump is not in use, click the pump into the Charger and
F
Note:
plug the system into an AC wall outlet (if possible) to charge
battery.
After the ³ end battery" signal has been activated or following
F
Note:
long periods of storage, wait 2 minutes after the pump has been
connected to an AC power supply before operating.
VET-PRO Operator's Manual
The default setting for a maximal rate is 500 ml/h. " " will
be shown on the right side of the display, indicating that the
program is in the third lock level only when the max. rate
will be set to a rate different than 500 ml/h.
§ The selected Locking Option will remain in the
pump's memory until changed by the user.
§ Whenever the pump is in more than one locking
level, the higher level will be displayed on the right
side of the screen.
Don't operate the pump on AC power if the battery is not
loaded in the pump for back up.
Bolus Rate
=> Maximal Rate
KVO Rate
Maximal Rate
500 ml/h
Adjust & press OK
53