5.
The pump will calculate all the set data and
display the Maximal rate for the program (to
be infused during Flat period). Press the
START/OK key to start the program.
Setting Program Delay– Refer to page 48.
F
During the stop or data setting mode, use the STOP/NO key to
Note:
return one step backwards: First press will erase the last digit
displayed, once all the digits are erased pressing the STOP/NO
key will display the previous screen .
6.
During the TPN program operation, the pump
will display the actual rate and program
direction (Up/Flat/ Down).
Program operation will start from 2 ml/h and
will increase to the maximal rate.
. During the Flat session the rate will not change
and the pump will show the following message:
. During the last part of the program (T3), the
program will start to decrease the infusion rate
according to the set taper down time.
F
If the program must be interrupted
Note:
early, and the planned taper down
time must commence immediately
press the STOP/NO key continually
(2 sec) to stop
and display Taper Down screen.
Set a new time or confirm the default setting. Press the START/OK
key to start the taper down process immediately.
34
the pump operation
T
Volume 1256 ml
P
Max Rate: 108 ml/h
N
Press OK to start
U
Rate
P
2
F
Rate
L
A
108
T
D
Rate
O
W
25
N
T
Taper Down: 0:30
P
Press OK key
N
VET-PRO Operator's Manual
ml/h
ml/h
ml/h