Introduction

This Manual Provides directions for the setup and operation of the VET-PRO
Pump
VET-PRO Features
S
Small, light and compact pump.
S
Multi Programs:
CONTINUOUS
25 STEPS
INTERMITTENT
S
Highly accurate fluid delivery. The piston mechanism ensures an accurate
volume delivery over more than 24 hours using the same administration set
segment
S
Operated with a customized safe administration set
S
Rechargeable internal battery
S
Can be latched into a Charger, which is mounted on an IV pole
S
A.B.S. Anti-Bolus System
S
Silent operation
4
Continuous program.
Infusion Rate
Volume: 0.1 to 9999 ml.
User can design a specific protocol, for drug
delivery, which contains up to 25 steps.
Infusion Rate: 0.1 to 1200 ml/hr.
Volume: up to 9999ml., For each step.
Accumulated Volume for all steps – up to 10
liters.
Rate Taper program, with programmable up
TPN
and down times
A set Dose delivered in set intervals. Between
Intervals the pump will allow a base rate or
KVO.
Patient Control Analgesia, continuous delivery
PCA
plus /or programmed boluses.
Bolus Rate: 0.1-1000 ml/h
0.1 to 1200 ml/hr.
:
VET-PRO Operator's Manual