VET-PRO VIP 2000 Operator's Manual page 38

Veterinary infusion pump
Table of Contents

Advertisement

Pump displays program name and status. The
status is either " Program ", "PCA Mode" or
"Epidural Mode".
The PCA program operates only in a lock
mode.
Program mode will be displayed if no protocol
has been locked.
PCA mode will be displayed if a protocol has
been locked and Epidural mode is Off.
Epidural mode will be displayed if a protocol
has been locked and Epidural mode is On – see
epidural mode chapter.
In all cases entering the PCA parameters and/or
changing them, requires a code.
Program Setting
F
Note:
To enter PCA menu, pump will require a code.
1.
Use numeric keys to set code 007 and confirm
with START/OK key.
Pump will display the PCA menu. Use the
.
arrow keys to scroll through the different
options, and select the required option, using
the START/OK key. Selecting the "More.."
option will display the Adjust Setting second
selection menu.
3.
Flushing is disabled during operation, therfore, always flush the set
before starting a program.
Selecting the flush option from the PCA menu, will display the flush
previous screen. Pressing the START/OK key will activate the flush
procedure. If no key will be pressed within 10 sec. the pump will return to the
PCA menu screen.
When the flush is completed the pump will go back to the PCA menu screen.
F
Note:
38
Whenever you decide to set a new program, follow
instructions at the bottom of the screen.
The flushing rate can be adjusted, in adjust setting options,
up to 1200 ml/h. The default is 600 ml/hr. During flushing
the air in line alarm is disabled.
P
PCA
C
A
PROGRAM
P
PCA
C
A
PCA MODE
P
PCA
C
A
Epidural Mode
P
Enter Code
C
007
A
Press OK key
P
=> Flush
C
MD Bolus
A
History
VET-PRO Operator's Manual