Choose "Adjust Setting" with START/OK key.
Use ARROW keys to scroll to "More" and
confirm with START/OK key.
Set technician Code (Code 902), and confirm
with START/OK key.
Use one of ARROW keys to change from OFF
to ON. Confirm with START/OK key.
The default setting is "Lock Out - NO".
Unlock in the same procedure.
"2" will be shown on the right side of the display, indicating that the
program is second level locked.
Level 3 – maximal rate locking
The operator can adjust the Maximal rate acceptable according to patient
needs:
Press info key twice.
Use ARROW keys to scroll to "Adjust Setting"
option.
Enter "Adjust Setting" pressing START/OK key.
Use ARROW keys to scroll to "More".
Confirm selection with START/OK key.
Set technician Code (Code 901), using the
numeric keys,
and confirm with START/OK
key.
52
Buzzer level
=> More ...
Show Setting
Technician
Code
Lock Out
ON
Adjust ?? & press OK
Volume Infused
Show Setting
=> Adjust Setting
Buzzer level
=> More ...
Adjust Setting
Technician
Code
VET-PRO Operator's Manual
902
901