Air In Line Alarm Test - VET-PRO VIP 2000 Operator's Manual

Veterinary infusion pump
Table of Contents

Advertisement

Air in Line - Alarm Test:
1.
Lift the door latch to open the pump door. Insert an administration set in the
tubing guide of the pump (as described on page 17).
2.
Close the door. Press the door until a click is heard.
3.
Set the infusion rate at 500ml/hr, using the
Numeric keyboard.
4.
Set Volume at 20 ml.
5.
Press the START/OK key to start operation. Turn the drip chamber upside
down, to allow an air bubble greater than 5 mm to enter into the
administration set.
6.
AIR alarm will be activated as soon as the air
bubble enters the set segment located behind
the pump door.
7.
Press the STOP/NO key to turn the alarm off.
8.
STOP appears on the display.
20
Rate
500 ml/h
Press OK for Volume
Volume
Press OK to start
Air in Line
Clean air
STOP
Press OK to start
VET-PRO Operator's Manual
20 ml
2 min

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents