Keysight N7020A User Manual page 12

Power rail probe
Table of Contents

Advertisement

1
General Information
Replacing Rigid, Solid Power Rail Probe Browser Tip, and Spring-Loaded
Power Rail Probe Browser Tips
The solid tips and spring-loaded tips are replaceable.
Spring loaded power rail probe browser tips offer a
method of probing signals that is less susceptible to
vibration or movement than traditional rigid tips.
Many users find it easier to use this type of tip. The
spring loaded tips work when they are either partially
or fully compressed and are protected against over compression damage.
To change the N7023A power rail probe browser tip,
use pliers to grip the tip and pull it straight out of its
contact socket along the axis of the N7023A power
rail probe browser. Do not grip the plastic insulator or
the housing with the pliers because the tip could be
crushed (see
unable to remove the tip and/or damaging the
N7023A power rail probe browser. Once the N7023A power rail probe browser tip
is removed, the new tip can be inserted with pliers into the contact socket along
the axis of the N7023A power rail probe browser. In order to insert the N7023A
power rail probe browser tip completely into the housing, carefully press the
browser tip against a hard surface.
Figure 4
You should exercise caution when using these sharp browser tips to avoid personal
WARNING
injury.
12
Figure
4). This could result in being
Proper Tip Removal Technique
N7020A User's Guide