6 Performance Verification

Procedure
Performance Test Record
This chapter tests that the probe meets its specified input resistance, which is
50 kΩ±2%.
Electrostatic discharge (ESD) can quickly and imperceptibly damage or destroy high
CAUTION
performance probes, resulting in costly repairs. Always wear a wrist strap when handling
probe components and ensure that cables are discharged before being connected.
Allow the probe to warm up for at least 20 minutes.
NOTE
Table 9
Required Test Equipment
Test Equipment
Oscilloscope
Digital Multimeter
Adapter
50
51
Critical Specification
Supported Oscilloscope
2 wire resistance accuracy >
BNC (f) to SMA (m) (In E2655C Kit)
Model Number
S-Series or 9000A series
Infiniium Oscilloscope
±
0.01%
34401A or equivalent
E2655-83201
49