Procedure - Keysight N7020A User Manual

Power rail probe
Table of Contents

Advertisement

6
Performance Verification

Procedure

1
Connect the equipment as shown in
2
Power on the oscilloscope with the oscilloscope application maximized.
3
Power on the DMM and select the 2-wire Ohm display on the DMM.
Figure 18
4
Read the DMM display for the input resistance.
5
Record the result in
between 49 kΩ and 51 kΩ.
50
Test Setup
Table 10
on page 51. To pass this test the result should be
Figure
18.
N7020A User's Guide

Advertisement

Table of Contents
loading

Table of Contents