5.6
Sweep Width Settings
Procedure
Setting by Pressing the SPAN WL or SPAN FREQ Soft Key
5-20
The following are three ways of setting the sweep width.
• Press the SPAN WL or SPAN FREQ soft key.
• Press the START WL/STOP WL or START FREQ/STOP FREQ soft key.
• Use the one-action key (soft key).
The following explains these procedures.
1.
Press SPAN. The soft key menu for sweep width settings appears along with the
sweep width setting screen.
2.
Press SPAN WL for wavelength measurement or SPAN FREQ for frequency
measurement.
3.
Enter a sweep width using the rotary knob, arrow keys, or numeric key pad.
4.
Press nm/ENTER.
Note
•
It is not necessary to press nm/ENTER when using the rotary knob or arrow keys.
•
The set value is applied to the measurement conditions area.
•
When a setting is changed,
•
If a value outside the setting range is entered, the nearest permitted value is set.
For instructions on switching the wavelength and frequency displays, see section 5.2.
appears in the measurement conditions area.
IM 735302-01E