Chapter 5 Mesurement; Auto Measurement - YASKAWA AQ6370B User Manual

Optical spectrum analyzer
Table of Contents

Advertisement

Chapter 5
Mesurement
5.1

Auto Measurement

Procedure
Explanation
IM 735302-01E
This key automatically sets the optimal measurement conditions for the light source
being measured, and performs measurement.
1.
Press SWEEP. The soft key menu regarding sweeping appears.
2.
Press the AUTO soft key. The soft key display reverses, and auto measurement
executes.
Waveform display example
Screen after auto sweep start value
The wavelength range of input light that can be auto-measured is 1200–1670 nm.
The following four items are set automatically, after which measurement is performed.
• Center wavelength (CENTER)
• Sweep width (SPAN)
• Reference level (REF LEVEL)
• Resolution (RESOLUTION)
After performing one auto sweep and setting optimal measurement conditions, a repeat
sweep is made for measurement.
During the automatic setting, only the following keys are enabled: REPEAT, SINGLE,
STOP, and UNDO/LOCAL (when under remote control).
Screen after auto sweep stop
1
2
3
4
5
6
7
8
9
10
11
App
Index
5-1