Editing The Grid Table - YASKAWA AQ6370B User Manual

Optical spectrum analyzer
Table of Contents

Advertisement

7.10 Editing the Grid Table

Procedure
Editing a Standard Grid Table
IM 735302-01E
1.
Press SYSTEM.
2.
Press the GRID EDITOR soft key. The Grid table edit screen appears.
Setting the Frequency Spacing
3.
Press one of the keys from 200 GHz SPACING to 12.5 GHz SPACING.
Depending on the selected soft key, a frequency spacing table of 200 GHz, 100
GHz, 50 GHz, 25 GHz, or 12.5 GHz can be used.
Setting the Reference Wavelength
4.
Press the REFERENCE WAVELENGTH soft key. A screen for specifying the
standard wavelength is displayed.
5.
Enter the standard wavelength using the rotary knob or arrow keys, then press
nm/ENTER.
1
2
3
4
5
6
7
8
9
10
11
App
Index
7-35