6.12 Searching

Procedure
Finding the Peak Wavelength/Level
Finding the Bottom Wavelength/Level
Finding the Next Peak/Bottom Level
6-42
1.
Press PEAK SEARCH. The menu for searching for the peak value is displayed.
2.
Press the PEAK SEARCH soft key. The moving marker is set on the waveform
peak (the maximum level value) and the marker value is displayed in the data
area.
2.
Continuing on from step 1, press the BOTTOM SEARCH soft key. The moving
marker is set on the waveform bottom (the minimum level value) and the marker
value is displayed in the data area.
Note
If the active trace is not set to DISP, the moving marker cannot be used. Set the trace VIEW
@ DISP/BLANK soft key setting to DISP.
Even if you press PEAK SEARCH the moving marker is displayed.
3.
With the moving marker displayed at a waveform peak or bottom, press the
NEXT LEVEL SEARCH soft key. The moving marker is placed on the next peak
(maximum level value) or bottom (minimum level value).
Finding the Level Peak/Bottom to the Right of the Marker
3.
With the moving marker displayed at a waveform peak or bottom, press the NEXT
SEARCH RIGHT soft key. The moving marker is placed on the peak (maximum
level value) or bottom (minimum level value) to the right of its current position.
Finding the Level Peak/Bottom to the Left of the Marker
3.
With the moving marker displayed at a waveform peak or bottom, press the NEXT
SEARCH LEFT soft key. The moving marker is placed on the peak (maximum
level value) or bottom (minimum level value) to the left of its current position.