Normalized Display - YASKAWA AQ6370B User Manual

Optical spectrum analyzer
Table of Contents

Advertisement

6.6

Normalized Display

Procedure
6-18
1.
Press TRACE. The soft key menu for trace settings appears.
2.
Press the ACTIVE TRACE soft key. Traces A through G appear in soft keys.
3.
Press the Trace G soft key. The Trace G setting menu is displayed.
4.
Press the VIEW G DISP/BLANK soft key and select DISP. Each time you press
the soft key DISP switches to BLANK, or viceversa.
5.
Press the CALCULATE G soft key.
6.
Press the NORMALIZE soft key.
7.
To normalize Trace A press the G=NORM A soft key, to normalize Trace B press
G=NORM B, and to normalize Trace G, press G=NORM C.
Example of waveform before normalization
Example of waveform after normalization
Note
Normalization is not possible if the waveform level is too low.
See section 6.5 for information about NORMALIZE.