Data Initialization - YASKAWA AQ6370B User Manual

Optical spectrum analyzer
Table of Contents

Advertisement

9.2

Data Initialization

Procedure
Explanation
9-2
1.
Press SYSTEM. The soft key menu regarding the system appears.
2.
Press the MORE 1/4 soft key to display 4/4.
3.
Press the PARAMETER INITIALIZE soft key. The initialization item setting menu
is displayed.
4.
Set the initialization type according to the items to be initialized.
5.
Press the EXECUTE soft key. The initialization executes.
To cancel, press the CANCEL soft key.
You can restore all settings to their factory defaults.
The following four types of initialization are available depending on the items to be
initialized.
• PARAMETER CLEAR
The parameter setting value of each function is initialized.
Waveform data of TRACE A - G is also initialized.
Use this to return the settings of the instrument to the already-known states.
• PARAM&DATA CLEAR
Parameter setting values as well as data including MEMORY and PROGRAM are
initialized.
• CAL DATA CLEAR
Alignment adjustment values and wavelength calibration values are initialized.
• ALL CLEAR
Current parameter setting values and data as well as alignment adjustment values
and wavelength calibration values are initialized.
IM 735302-01E