7.14 Correcting Displayed Values

Procedure
Setting the Wavelength Shift Amount
Setting the Level Shift Amount
IM 735302-01E
1.
Press SYSTEM.
2.
Press the WL SHIFT soft key. The wavelength shift setting screen is displayed.
3.
Enter a wavelength shift amount using the rotary knob, arrow keys, or numeric
key pad.
4.
Press ENTER.
1.
Press SYSTEM.
2.
Press the LEVEL SHIFT soft key. The level shift setting screen is displayed.
3.
Enter a level shift amount using the rotary knob, arrow keys, or numeric key pad.
4.
Press ENTER.
Note
After entering the wavelength or level shift amount, the setting is applied to the displayed
values upon the next measurement.