Attaching the Connector Adapter

Attachment Procedure
Removal Procedure
IM 735302-01E
Attach the optional connector adapter before using the instrument.
Always turn the power OFF before replacing the connector adapter. The instrument
has a built-in reference light source for wavelength calibration, and infrared light is
always being output from the optical output connector. Never look into the optical
output connector. Infrared light entering the eyes can cause severe injury and loss
of vision.
Note
A different connector adapter is used for OPTICAL INPUT and CALIBRATION OUTPUT. Make
sure not to use the wrong connector adapter.
1.
Confirm that the power is OFF.
2.
Open the optical connector cover at the front of the instrument.
3.
Clean the ferrule edge of the optical I/O section using a swab soaked with a small
amount of pure alcohol.
4.
Insert the connector adapter all the way in.
5.
Push the connector adapter's lock lever down.
The adapter has been attached correctly if the groove in the lock lever interlocks with the
latch pin of the optical input/output section.
1.
Confirm that the power is OFF.
2.
Turn the connector adapter's lock lever up. The lock lever's lock is released.
3.
Pull the connector adapter all the way out.
4.
Close the optical connector cover at the front of the instrument.
Optical input
section
Ferrule
Latch pin
Panel
WARNING
Connector
adapter
Lock lever
Attached
3-3
1
2
3
4
5
6
7
8
9
10
11
App
Index