5.16 Trigger Output

Procedure
Setting the Trigger Output Mode
Explanation
5-42
1.
Press SYSTEM. The soft key menu for the system settings appears.
2.
Press the MORE soft key. The MORE 2/4 soft key menu is displayed.
3.
Press the TRIG OUTPUT MODE soft key. The trigger output setting menu is
displayed.
4.
Press the SWEEP STATUS soft key.
Note
When synchronous sweep measurement is specified, the TRIG OUTPUT MODE soft key is
disabled. To enable it, turn TLS SYNC SWEEP OFF. For the procedure, see section 5.17,
"Synchronous Sweep Measurement with a Turnable Laser Source."
When pulse light measurement is specified, external trigger output cannot be performed.
Turn OFF PLS LIGHT MEASURE.
You can synchronize sweeping from the trigger output terminal on the rear panel of the
instrument and output a positive logic signal. Signals are only output when sweeping.
Output signals are of TTL levels.