Synchronous Sweep Measurement With The Tunable Laser Source - YASKAWA AQ6370B User Manual

Optical spectrum analyzer
Table of Contents

Advertisement

5.17 Synchronous Sweep Measurement with the
Tunable Laser Source
Procedure
Connecting to the Tunable Laser Source
IM 735302-01E
Before turning on the power supply, make the following connections.
MAIN POWER
ON
OFF
100-240V AC
150VA MAX 50/60Hz
FUSE 250V T 5A
AQ6370B
Connecting the Turnable Laser Source When Using the AQ4321 Series
1.
Connect the GP-IB2 connector of the instrument and the GP-IB connector of the
AQ4321A/D using the GP-IB cable.
2.
Connect the TRIG IN terminal of the instrument and the SMPL TRIG OUT terminal
of the AQ4321A/D using a coaxial cable.
3.
Connect the TRIG OUT terminal of the instrument and the OSA SYNC IN terminal
of the AQ4321A/D using a coaxial cable.
Connecting the Turnable Laser Source When Using the AQ2200-136
1.
Connect the GP-IB2 connector of the instrument and the GP-IB connector of the
AQ2201/2202 mainframe using the GP-IB cable.
2.
Connect the TRIG IN terminal of the instrument and the SMPL TRIG OUT terminal
of the AQ2201/2202 mainframe using a coaxial cable.
3.
Connect the TRIG OUT terminal of the instrument and the TRIGGER IN terminal
of the AQ2201/2202 mainframe using a coaxial cable.
Coaxial cable
Coaxial cable
TRIGGER
TRIGGER
ANALOG
IN
OUT
OUT
SERIAL
ETHERNET
VIDEO OUT
KBD
(RS-232)
10/100BASE-TX
(SVGA)
GP-IB1
(IEEE488.1/488.2)
GP-IB2
(IEEE488.1)
(FOR TLS,ETC.)
GP-IB cable
WARNING
Do not operate without reading the
safety precautions in the user' s manual.
TRIGGER IN
TRIGGER OUT
GP-IB
Tunable Laser Source
5-43
1
2
3
4
5
6
7
8
9
10
11
App
Index

Advertisement

Table of Contents
loading

Table of Contents