5.15 External Trigger Measurement
Procedure
Setting the Trigger Input Mode
IM 735302-01E
1.
Press SYSTEM. The soft key menu regarding the system settings appears.
2.
Press the MORE soft key. The MORE 2/4 soft key menu is displayed.
3.
Press the TRIG INPUT MODE soft key. The trigger input mode's setting menu is
displayed.
4.
To set sample trigger mode or sweep trigger mode, press the SMPL TRIG or
SWEEP TRIG soft key, respectively.
If you set SWEEP TRIG, that concludes the settings.
If you press the EXEC soft key, the instrument returns to the original screen.
Note
•
When using a sample trigger, turn TLS SYNC OFF.
•
When using a sweep trigger, turn TLS SYNC OFF. Also, TRIGGER SETTING and EXT
TRIGGER MODE cannot be set.