Displaying Calculated Waveforms - YASKAWA AQ6370B User Manual

Optical spectrum analyzer
Table of Contents

Advertisement

6.5

Displaying Calculated Waveforms

Procedure
Selecting Traces for Calculation
Selecting a Calculation
IM 735302-01E
1.
Press TRACE. The soft key menu for trace settings appears.
2.
Press the ACTIVE TRACE soft key. Traces A through G appear in soft keys.
3.
Press the soft key corresponding to a trace that is available for trace-to-trace
calculation (C, F, or G).
The calculated results are displayed in the selected trace.
The items that can be calculated differ depending on the trace selected.
4.
Press the VIEW soft key of the selected trace and select DISP. Each time you
press the soft key DISP switches to BLANK, or viceversa.
5.
Press the CALCULATE soft key. The LOG and Linear selection menus are
displayed.
6.
For log calculations, press the LOG MATH soft key. For linear calculations, press
the LIN MATH soft key. The soft key menu showing the choices for calculation
appears.
7.
Press the soft key corresponding to the desired calculation. The calculation
executes.
Note
Trace-to-trace calculations can only be performed on C, F, or G. If the active trace is set to
something other than trace C, F, or G, the CALCULATE key is disabled.
If a trace targeted for calculation is remeasured and the center wavelength and
measurement span of the trace being measured is changed, it is recalculated and
redisplayed.
If the measurement condition (resolution) of the trace targeted for calculation does not
match, a warning is displayed after calculation.