LifeVest System WCD 3100 Training Manual
When the test is finished
Page 5-6
7
Press the response buttons to start the test.
8
A message tells you the time remaining to finish the test.
The monitor sounds a loud alarm near the end of the test.
If you want to stop the test before completion, press both
response buttons.
If you get an error message during the test, see next
page.
1
When the test is finished, one of the following messages
displays:
2
Disconnect the test plug from the monitor, then remove the
battery from the monitor
Use the monitor only if you receive the
PASS message.
If you receive the FAIL message, do not
use the monitor. The system becomes
inoperable. Contact ZOLL Lifecor for a
replacement.
ZOLL Lifecor internal use only