ZOLL WCD 3100 Training Manual page 49

Lifevest system
Table of Contents

Advertisement

Treatment menu and settings
Menu
Menu option
VT/VF rate threshold
VT/VF response time
MD notification option
Patient sleep interval
Response time
extension
Pulse energy #1...5
Post treatment
message
ZOLL Lifecor internal use only
What it means
Heart rate that must be sustained
before VT or VF is declared.
Elapsed time before treatment
delivered.
Determines if patient receives "call
doctor" message after receiving a
treatment.
Time the patient normally goes to
sleep and awakens.
Time to be added to the response
time during the patient sleep
interval.
Energy level of each of the five
shocks.
Message displayed after a
treatment shock.
Advanced programming
How to select
Range: 120-250 beats per minute
Defaults: VT=150 beats per
minute, VF=200 beats per minute
Range: VT=60-180 seconds,
VF=25-55 seconds
Defaults: VT=60 seconds, VF=25
seconds
Default: Off
Can be set to repeat message
every 5 minutes for up to 1 hour
Range: Any two times during a
24-hour clock
Default: Asleep at 00:00 (midnight)
and awake at 06:00 (6:00 a.m.)
Range: 0-30 seconds
Default: 0
Range: 75-150 joules
Default: 150 joules
Optional. Can be set to any
alphanumeric message you want
to display after the patient receives
treatment. For example, can be set
to display "CALL 911" or other
phone number for the patient to
call.
Page B-11