ZOLL WCD 3100 Training Manual page 31

Lifevest system
Table of Contents

Advertisement

If you have a problem testing the system
If you have a problem testing the system, use this chart to determine how to
correct the problem.
If you have any other problem testing the system, call ZOLL Lifecor.
Message
ZOLL Lifecor internal use only
What it means
System has a problem that requires
servicing. The monitor is inoperable
and cannot be used.
Battery doesn't have enough
reserve power to run test.
The test was stopped, probably by
pressing the response buttons. The
monitor may also be in training
mode.
Other procedures
What to do
Send data via modem. Contact
ZOLL Lifecor to report problem and
arrange to have monitor replaced.
Remove battery and put a fully-
charged battery into the monitor.
Start the test again.
Remove the battery, put it back into
the monitor, then start the test
again. Don't press any buttons
during the test. Make sure the
monitor is not in training mode.
Page 5-7