Overview Of Signal Paths For Camera Control - Sirona Intego Service Manual

Hide thumbs Also See for Intego:
Table of Contents

Advertisement

5 Important information
5.9 Overview of signal paths for camera control
32
5.9

Overview of signal paths for camera control

The presence of the camera SiroCam F / AF is sent to DAB via Sivision
DARWIN
connect.
● Camera in instrument holder of
TS dentist element/CS dentist
element (holder 5)
A photoelectric barrier enables the
DAB board to detect the removal of
the camera.
● The removal signal for the camera is transferred to SIVISION
connect on the external PC via CAN > Ethernet.
● SIVISION connect sends the camera ON signal to the camera via
USB.
● The user interface switches over to the PC dialog box when all other
instruments have been set down.
● Sidexis/SiVideo represents the live image of the camera in video
overlay, even when a different instrument is active.
● If the camera is the active instrument, the camera functions are
controlled via the connected foot switch.
5.10
Acoustic feedback from the treatment center
If several problems occur simultaneously, the alarm with the highest
priority will always sound. The alarm with the lower priority will only sound
automatically once this problem has been eliminated.
Sirona Dental Systems GmbH
Service Manual INTEGO / INTEGO pro
● Camera in TS external
instrument holder (holder 6)
● Camera in compact/comfort
assistant element (holder 2)
The camera detects its own
removal thanks to magnets inside
the holder. It sends this information
via USB > SIVISION connect >
Ethernet > DSL > DSA to the DAB
64 57 316 D3543
D3543.076.01.04.02 06.2015

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Intego pro

Table of Contents