Sirona Dental Systems GmbH
Service Manual INTEGO / INTEGO pro
LED
3V3
RUN
ERR
TX
RX
ANY KEY PRESSED
64 57 316 D3543
D3543.076.01.04.02 06.2015
6.3.3

Board DAT - EasyPad

Components on the board
LEDs
Color
Function
Green
Generated +3.3 V voltage
Green
Flashes 1x per second during normal operation
Red
Flashing:
● 2x Loss of communication with DSA
● 3x Serious hardware error (CRC error)
● 4x Serious hardware error (shift register error)
yellow
Illuminated: Data transmission active
Flashing: Cyclic if there is no communication
(Two beeps emitted if no connection can be established when the
button is pressed)
yellow
Illuminated: Data reception active
yellow
Illuminated: Any key pressed