Inlet and outlet run
If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc.
Compliance with the following inlet and outlet runs is recommended in order to ensure
measuring accuracy.
Fig. 2: Inlet and outlet run
1
Valve (2/3 open)
2
Pump
3
Two pipe bends in different directions
Endress+Hauser
³ 15 x DN
³ 20 x DN
1
³ 20 x DN
2
³ 15 x DN
3
³ 3 x DN
Installation