Endress+Hauser Proline Prosonic Flow 93T Portable Operating Instructions Manual page 16

Portable ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93T Portable:
Table of Contents

Advertisement

Installation
2.6.4 Mounting with strapping bands (flexible)
For sensor Prosonic Flow P (DN 50 to 4000 / 2 to 160")
"
Caution!
• Each time you use the strapping bands, check that the ratchet locks and springs function safely
beforehand.
• Inspect the strapping bands for damage.
Procedure
Closing the strapping band lock
1.
Fit the mounting bolt onto the strapping band.
2.
Guide the strapping band around the pipe making sure it is not twisted in the process and,
with the ratchet lock (a) open, push the end through the slot. Pretension manually by
pulling on the free end of the strapping band.
!
Note!
If you do not pretension the bands it is more difficult to release the strapping bands.
3.
Tension continuously by moving the lever back and forth (b) until the strapping band is
optimally tensioned.
4.
Then push down the lever (c).
"
Caution!
The tensioning clamp (d) must engage on both sides!
Opening the strapping band lock
1.
Pull back the lever lock (e) while simultaneously opening the lever 180° (f) until the lever
lock (g) is engaged.
2.
Remove the strapping band.
a
Fig. 10: Strapping band lock
1
Closing the strapping band lock
2
Opening the strapping band lock
16
c
b
d
1
Proline Prosonic Flow 93T Portable
f e
2
g
g
A0011556
Endress+Hauser

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents