Proline Prosonic Flow 93T Portable
2.6
Mechanical preparation
The way in which the sensors are secured differs on account of the pipe nominal diameter and
the sensor type.
Overview of possible ways to secure the various sensors:
Sensor
For the measuring range
P
DN 15 to 65 (½ to 2½")
P
DN 50 to 4000 (2 to 160")
2.6.1 Mounting the sensor holder
Sensor: Prosonic Flow P (DN 15 to 65 / ½ to 2½")
1.
Set the sensor distance determined (e.g. C9) on the sensor holder.
– Position the sensor holders with the aid of the mounting rail.
27. 25 . 23
. 21. 19 . 17 . 15 . 13 . 11 . 9 . 7 . 5 . 3 . 1
27 25 23
Fig. 4: Setting the sensor distance with the aid of the mounting rail
A
Sensor distance for measurement via one traverse
B
Sensor distance for measurement via two traverses
Endress+Hauser
Pipe nominal diameter
DN 15 to 65 (½ to 2½")
DN ≤ 200 (8")
DN > 200 (8")
DN 50 to 4000 (2 to 160")
A
K I H G F E
D C B A
K I H G F E
Secured by
Sensor holder
Strapping bands
(metal, medium nominal diameters)
Strapping bands
(metal, large nominal diameters)
Mounting with strapping bands
(flexible)
27. 25 . 23 . 21. 19 . 17 . 15 . 13 . 11 . 9 . 7 . 5 . 3 . 1
K I H G F E D C B A
Installation
→ ä 17
→ ä 14
→ ä 15
→ ä 16
B
A0013546
11