6 Maintenance
6.2 Inspection and maintenance
88
6.2

Inspection and maintenance

Inspection and preventive maintenance must be performed at scheduled
intervals to protect the health and safety of patients, users, and other
persons.
The information provided in the document
and safety-related checks"
document can be downloaded at http://www.sirona.com/de.html.
Annual inspection
In order to ensure the operational safety and functional reliability of your
product, you as the system owner should check the equipment at regular
intervals (at least once a year) or commission your dental depot to do so.
Maintenance by the service engineer
In addition to the annual check to be carried out by the system owner or
authorized persons, preventive maintenance must be performed after 4,
7 and 10 years, and then at two-year intervals.
Checking image quality
The image quality should be assessed by the system owner at regular
intervals, at least once a year.
On digital image receptor systems, the degree of postprocessing
(brightness or contrast adjustment) that is required in the image
processing software (e.g. SIDEXIS) to produce satisfactory results is
used as an assessment criterion.
Checking image quality
The image quality should be assessed by the system owner at regular
intervals, at least once a year. The constancy test is described in REF 63
36 924
On digital image receptor systems, the degree of postprocessing
(brightness or contrast adjustment) that is required in the image
processing software (e.g. SIDEXIS) to produce satisfactory results is
used as an assessment criterion.
If, after taking into account the patient's anatomy and excluding possible
sources of error such as incorrect patient positioning, this criterion seems
to apply, immediately contact a service engineer to have potential system
faults repaired.
Country-specific requirements
Observe any possible additional country-specific requirements.
Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS XG 5 / Ceph
"Inspection and maintenance
" 59 87 685"
REF
should be helpful here. The
D3352.201.04.25.23 12.2017
60 04 811 D3352