Repair and Replacement Procedures
Removing and Replacing the A13 Frequency Offset Receiver Assembly
Removing and Replacing the A13 Frequency Offset
Receiver Assembly
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16-inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-9
for this procedure.
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to
3. With a 5/16-inch torque wrench, disconnect the three semirigid cables (item
flexible cable (item
Be careful not to damage the center pins of the semirigid cables. Some flexing
CAUTION
of the cables is necessary to remove the assembly. Do not over-bend the
semirigid cables.
4. Disengage the A13 frequency offset receiver assembly from the A11 connector and slide
out of the analyzer. Observe the
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
7-24
②
) from the top of the A13 frequency offset receiver assembly.
CAUTION
Table 7-2 on page
PNA Series Microwave Network Analyzers
"Removing the Covers" on page
above.
7-80.
E8362B, E8363B, E8364B
7-6.
①
) and
Service Guide E8364-90026