Power Supply Check - Agilent Technologies E8362B Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B

Power Supply Check

There are no fuses to replace within the power supply. If you determine that
NOTE
the power supply is the failed assembly, replace the power supply.
A catastrophic failure in the power supply can be determined by observing the line switch,
and by measuring the probe power at the front-panel Probe Power connectors:
1. Ensure that the instrument is plugged in with the power switch in the standby
position (power not switched on). Verify that the line switch glows yellow. A line
switch that glows yellow indicates that the +15 VDC line (P15 STB) is providing
enough voltage to light the LED. (The actual voltage may not be +15 VDC.)
2. Turn on the instrument power and verify that the line switch glows. When the line
switch glows, it is an indication that the power supply has received an "ON"
command and that the +5.2 VDC supply can supply enough current to light, at least,
these lamps.
3. The front-panel probe power-connector can be used to check the +15 VDC and
−12.5 VDC (−15 VDC) supplies. The −12.5 VDC is produced by post regulating the
−15 VDC supply. Refer to
Figure 4-2 Probe Power Connector
If all of these supply voltages are missing, it is likely that the problem is either a defective
A4 power supply, or another assembly is loading down the A4 power supply. Continue with
"If All Supply Voltages Are Missing" on page
If the line switch is lit correctly, and the probe power voltages measure within the
specifications listed in
catastrophic failure; however, the power supply could still be at fault. Continue with the
next section to measure the individual voltage supplies.
Service Guide E8364-90026
Figure 4-2
for a diagram of the probe power connector.
PROBE POWER
+15 V
Table 4-1 on page
4-9, the power supply has not suffered a
-12.6 V
sa869a
4-10, to determine the cause of the problem.
Troubleshooting
Power Up Troubleshooting
4-7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8363bE8364b