Tests and Adjustments

Option H11, IF Access, Operation Verification

Option H11, IF Access, Operation Verification
To verify that the Option H11, IF Access, is functioning properly:
1. Download and install the verification program from the Web:
a. Go to website: http://na.tm.agilent.com/pna/support/.
b. Download file: H11VerifyInstall.exe to the C:\ drive on the PNA.
c. After the file has finished loading, double click on the filename. At this point the file
will self-extract and load into its default file location. Allow this process to complete,
uninterrupted.
2. Create (if necessary) a macro for the verification program:
a. On the analyzer's System menu, point to Macro. If a macro named H11Verify is
present in the list, proceed to step 3. If not, continue here to create the macro.
b. Click Macro Setup.
c. In the Macro Setup window, click on the first blank line in the list (the next line
after the last entry). Click Edit.
d. In the Edit Macro Setup window, enter H11Verify as the Macro Title.
e. Click Browse.
f. In the Open window, navigate to Network Analyzer/Service/H11Verify.exe. Click
Open. This places the H11Verify.exe filename in the Macro Executable box in the
Macro Setup window.
g. In the Macro Setup window, click OK.
h. The macro should now be entered in the list in the Macro Setup window.
i. Click OK.
3. Run the H11Verify macro:
a. On the analyzer's System, Macro menu, click H11Verify.