Tests and Adjustments
System Verification
Precautions for Handling Airlines
The verification kit includes airlines. Follow the steps below to ensure good measurement
techniques.
• Be very careful not to drop the airline's center or outer conductor. Irreparable damage
will result if these devices are dropped.
• During this procedure, you will be touching the exposed center conductor of the test
port and the center conductor of the airline.
— Observe proper ESD procedures.
— Clean your hands or wear gloves; skin oils will cause a change in electrical
performance.
• Refer to
Figure
3-9,
procedures.
Figure 3-9
Aligning the Center Conductor
Figure 3-10 Torquing the Connection
3-24
Figure
3-10, and to your calibration kit manual for connection
PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
Service Guide E8364-90026