Option 016, Receiver Attenuators; Options 040 And 041, Extended Frequency Range To 40 Ghz; Options 050 And 051, Extended Frequency Range To 50 Ghz; Option 080, Frequency Offset Mode - Agilent Technologies E8362B Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

General Product Information
PNA Series Microwave Network Analyzers
Analyzer Options Available
E8362B, E8363B, E8364B
For high dynamic range measurements, front panel jumpers are moved to reverse the
signal path through one of the couplers, allowing for a 15 dB improvement in transmitted
signal sensitivity in one direction only. Jumpers are installed on both ports allowing the
user to choose a measurement in either the forward or reverse direction.

Option 016, Receiver Attenuators

This upgrade provides a 35-dB step attenuator, in 5-dB steps, in the A and B receiver
channels.

Options 040 and 041, Extended Frequency Range to 40 GHz

This upgrade—E8362BU-040—provides the additional hardware and software necessary
to make frequency measurements up to 40 GHz with your E8362B PNA.
If you have a standard E8362B PNA, then upgrade E8362BU-040 provides the additional
hardware and software necessary to make frequency measurements up to 40 GHz.
If your E8362B PNA already has option 014, then option E8362BU-041 provides additional
hardware and software necessary to make frequency measurements up to 40 GHz, and
upgrades your 3.5 mm connectors to 2.4 mm connectors.

Options 050 and 051, Extended Frequency Range to 50 GHz

This upgrade provides the additional hardware and software necessary to make frequency
measurements up to 50 GHz with your E8362B/63B PNA.
If you have a standard E8362B PNA, then upgrade E8362BU-050 provides the additional
hardware and software necessary to make frequency measurements up to 50 GHz.
If your E8362B PNA already has option 014, then option E8362BU-051 provides additional
hardware and software necessary to make frequency measurements up to 50 GHz, and
upgrades your 3.5 mm connectors to 2.4 mm connectors.
If you have a E8363B PNA—with or without option 014—then upgrade E8363BU-050
provides the additional hardware and software necessary to make frequency
measurements up to 50 GHz.
1
Option 067 and 068
, Extended Frequency Range to 67 GHz
These upgrades provide the additional hardware and software necessary to make
frequency measurements up to 67 GHz with your E8363B/64B PNA. Option 067 is for
analyzers without Option 014 and Option 068 is for analyzers with Option 014 already
installed.

Option 080, Frequency Offset Mode

This upgrade provides the additional hardware and software necessary to make frequency
offset measurements. This option requires that Option 081 also be installed.
1. These upgrades are not available for the E8362B PNA.
2-4
Service Guide E8364-90026