PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362B, E8363B, E8364B
Preliminary Checks
Figure 3-6
Typical Cable Magnitude and Phase Stability Response
If the Cable Magnitude and Phase Stability Check Fails
1. Clean the cable and devices and torque connections to specifications (refer to
"Review
the Principles of Connector Care" on page
3-5). Repeat the check.
2. If the check still fails, the cable should be repaired or replaced.
Service Guide E8364-90026