Quick Start Guide
E4 Certificate: CML17JPN1145X
Ex d [ia] IIC T6 Ga/Gb (integral transmitter and sensor)
Ex d [ia Ga] IIC T6 Gb (remote transmitter)
Ex ia IIC T6 Ga (remote sensor)
Ambient temperature range: -20 °C to +60 °C
Process temperature range: -202 °C to +427 °C
42 VDC Max, 4-20 mA HART
32 VDC Max, Fieldbus
U
= 250V
m
Special Conditions for Safe Use (X):
1. For information regarding the dimensions of the flameproof joints, the manufacturer
shall be contacted.
2. The Flowmeter shall be provided with special fasteners of property class A2-70 or
A4-70.
3. Units marked with "Warning: Electrostatic Charging Hazard" may use non-conductive
paint thicker that 0.2mm. Precaution shall be taken to avoid ignition due to
electrostatic charge on the enclosure.
5.6 Brazilian Certifications (INMETRO)
I. S. Certification
ABNT NBR IEC 60079-0: 2013
ABNT NBR IEC 60079-11: 2013
ABNT NBR IEC 60079-26: 2008 and 2009 correction
ABNT NBR IEC 60529: 2011
Portaria INMETRO no. 179: 18 May 2010
I2 Certification Number: TÜV 16.1958 X
Ex ia IIC T4 Ga (-60 °C  Ta  + 70 °C) 4-20 mA HART
Ex ia IIC T4 Ga (-60 °C  Ta  + 60 °C) Fieldbus
IB Certification Number: TÜV 16.1958 X
Ex ia IIC T4 Ga (-60 °C  Ta  + 60 °C)
Special Conditions for Safe Use (X):
1. When fitted with 90V transient suppressors, the equipment is not capable of passing
the 500V insulation test. This must be taken into account upon installation.
2. The enclosure may be made from aluminum alloy with a protective polyurethane paint
finish; however, care should be taken to protect it from impact or abrasion when
located in Zone 0. The polyurethane paint finish may constitute an electrostatic hazard
and must only be cleaned with a damp cloth.
3. When the equipment is installed, particular precautions must be taken to ensure,
taking into account the effect of process fluid temperature, that the ambient
temperature of the electrical housing of the equipment meets the marked protection
type temperature range.
Flameproof Certification
ABNT NBR IEC 60079-0: 2013
ABNT NBR IEC 60079-1: 2009 and 2011 correction
ABNT NBR IEC 60079-11: 2013
ABNT NBR IEC 60079-26: 2008 and 2009 correction
Portaria INMETRO no. 179: 18 May 2010
30
April 2016