Download Print this page

Emerson Rosemount 8800D Series Quick Start Manual page 23

Vortex flowmeter
Hide thumbs Also See for Rosemount 8800D Series:

Advertisement

April 2016
ATEX FISCO
IA Certification No. Baseefa05ATEX0084X
ATEX Marking
II 1 G Ex ia IIC T4 Ga (-60 °C  Ta  60 °C)
2460
Special Conditions for Safe Use (X):
1. When fitted with 90V transient suppressors (T1 option), the equipment is not capable
of passing the 500V isolation test. This must be taken into account upon installation.
2. The enclosure may be made from aluminum alloy and given a protective polyurethane
paint finish; however, care should be taken to protect it from impact or abrasion when
located in Zone 0 environment. The polyurethane paint finish may constitute an
electrostatic hazard and must only be cleaned with a damp cloth.
3. When the equipment is installed, particular precautions must be taken to ensure taking
into account the effect of process fluid temperature, that the ambient temperature of
the electrical housing of the equipment meets the marked protection type
temperature range.
ATEX Type n Certification
EN 60079-0: 2012 + A11: 2013
EN 60079-11: 2012
EN 60079-15: 2010
N1 Certification No. Baseefa05ATEX0085X
ATEX Marking
II 3 G Ex nA ic IIC T5 Gc (-50 °C  Ta  70 °C) 4-20 mA HART
II 3 G Ex nA ic IIC T5 Gc (-50 °C  Ta  60 °C) Fieldbus
Maximum Working Voltage = 42 VDC 4-20 mA HART
Maximum Working Voltage = 32 VDC Fieldbus
Special Conditions for Safe Use (X):
1. When fitted with 90V transient suppressors (T1 Option), the equipment is not capable
of passing the 500V isolation test. This must be taken into account upon installation.
2. The enclosure may be made from aluminum alloy with a protective polyurethane paint
finish. The polyurethane paint finish may constitute and electrostatic hazard and must
only be cleaned with a damp cloth.
3. When the equipment is installed, particular precautions must be taken to ensure,
taking into account the effect of process fluid temperature, that the ambient
temperature of the electrical housing of the equipment meets the marked protection
type temperature range.
ATEX Dust Certification
EN 60079-0: 2012 + A11: 2013
EN 60079-31: 2014
ND Certificate: BaseefaATEX17.0020X
II 2 D Ex tb IIIC T85°C Db (-20 °C ≤ Ta ≤ 70 °C)
2460
Maximum Working Voltage = 42 VDC 4-20 mA HART
Maximum Working Voltage = 32 VDC Fieldbus
Special Conditions for Safe Use (X):
1. The enclosure may be made from aluminium alloy with a protective polyurethane paint
finish which may constitute a potential electrostatic ignition risk. Care should be taken
to protect it from external conditions conducive to the build-up of electrostatic charge
Quick Start Guide
23

Advertisement

loading